摘要

In this study, conventional thin-film microextraction (TFME) was endowed with magnetic by introducing superparamagnetic SiO2@Fe3O4 nanoparticles in thin-films. Novel magnetic octadecylsilane (ODS)-polyacrylonitrile (PAN) thin-films were prepared by spraying, and used for the microextraction of quetiapine and clozapine in plasma and urine samples, followed by the detection of HPLC-UV. The influencing factors on the extraction efficiency of magnetic ODS-PAN TFME, including pH, extraction time, desorption solvent, desorption time, and ion strength were investigated systematically. Under the optimal conditions, both analytes showed good linearity over ranges of 0.070-9.000 mu g mL(-1) and 0.012-9.000 mu g mL(-1) in plasma and urine samples, respectively, with correlation coefficients (R-2) above 0.9990. Limits of detection (LODs) for quetiapine in plasma and urine samples were 0.013 and 0.003 mu g mL(-1), respectively. LODs for clozapine in plasma and urine samples were 0.015 and 0.003 mu g mL(-1), respectively. The relative standard deviations (RSDs) for quetiapine and clozapine were less than 9.23%. After the validation, the protocol was successfully applied for the determination of quetiapine and clozapine in patients' plasma and urine samples with satisfactory recoveries between 99-110%. The proposed magnetic ODS-PAN TFME was very simple, fast and easy to handle. It showed high potential as a powerful pretreatment technology for routine therapeutic drug monitoring (TDM) in plasma and urine samples.